Improving echo-planar imaging-based diffusion weighted imaging of the head and neck using local shim coils

Sven Stephan Walter, Joerg Rothard, Mike Notohamiprodjo, Miriam Keil, Petros Martirosian, Sergios Gatidis, Konstantin Nikolaou

Abstract


Purpose: To evaluate the performance of additional local shim coils within the head/neck surface coil for diffusion-weighted imaging (DWI) MRI of the head/neck region using a 3-Tesla scanner.

Material and Methods: Ten healthy volunteers underwent single-shot echo-planar DWI and T1 gradient echo acquisition of the head/neck region, without (sEPI; sGRE) and with (cEPI; cGRE) the use of additional shim coils integrated into the head/neck coil. Additionally, B0 field maps with and without additional shim coils were acquired using a dual-echo T1w GRE sequence. Acquisitions were performed on a clinical 3T scanner. In the DWI images, visual evaluation was performed for overall image quality (OAIQ), signal loss, distortions and ghosting. Failure of fat saturation was visually assessed for all sequences. Assessments were quantified using a 4-point scale by two radiologists in consensus. The effects on static field homogeneities were evaluated visually in the acquired field maps. Apparent diffusion coefficients (ADCs) were quantified in the spinal cord, spinal fluid and trapezius muscle for sEPI and cEPI.

Results: Compared to sEPI, cEPI significantly improved OAIQ, distortion and ghosting and decreased signal loss (p≤0.008). Additionally, cEPI resulted in significantly improved fat saturation compared to sEPI, sGRE and cGRE (p<0.001). Regarding the static magnetic field, additional local shim coils resulted in marked homogenisation, especially in the lower/posterior neck. ADC values did not differ significantly (p≥0.08).

Conclusions: Additional local shim coils integrated into the head/neck surface coil improve homogenisation of the static magnetic field and providing improved DWI quality compared to standard sEPI. For T1-weighted sGRE, additional coils have no significant effect on fat saturation.


Keywords


Magnetic resonance imaging; echo-planar imaging (EPI); diffusion-weighted imaging (DWI); head and neck; shimming

Full Text:

PDF

References


Bae YJ, Choi BS, Jeong HK, et al. Diffusion-weighted imaging of the head and neck: influence of fat-suppression technique and multishot 2D navigated interleaved acquisitions. AJNR Am J Neuroradiol 2018; 39(1): 145-150.

Hirata K, Nakaura T, Okuaki T. Comparison of the image quality of turbo spin echo- and echo-planar diffusion-weighted images of the oral cavity. Medicine (Baltimore) 2018; 97(19): e0447.

Mikayama R, Yabuuchi H, Sonoda S, et al. Comparison of intravoxel incoherent motion diffusion-weighted imaging between turbo spin-echo and echo-planar imaging of the head and neck. Eur Radiol 2018; 28(1): 316-324.

Chilla GS, Tan CH, Xu C. Diffusion weighted magnetic resonance imaging and its recent trend-a survey. Quant Imaging Med Surg 2015; 5(3): 407-422.

Schellinger PD, Bryan RN, Caplan LR. Evidence-based guideline: The role of diffusion and perfusion MRI for the diagnosis of acute ischemic stroke: report of the therapeutics and technology assessment subcommittee of the american academy of neurology. Neurology 2010; 75(2): 177-185.

Fung SH, Roccatagliata L, Gonzalez RG. MR diffusion imaging in ischemic stroke. Neuroimaging Clin N Am 2011; 21(2): 345-377, xi.

Maier SE, Sun Y, Mulkern RV. Diffusion imaging of brain tumours. NMR Biomed 2010; 23(7): 849-864.

Ma G, Xu XQ, Hu H, et al. Utility of readout-segmented echo-planar imaging-based diffusion kurtosis imaging for differentiating malignant from benign masses in head and neck region. Korean J Radiol 2018; 19(3): 443-451.

Walter SS, Liu W, Stemmer A, et al. Combination of integrated dynamic shimming and readout-segmented echo planar imaging for diffusion weighted MRI of the head and neck region at 3Tesla. Magn Reson Imaging 2017; 42: 32-36.

Jin J, Chen M, Li Y, et al. Detecting acute myocardial infarction by diffusion-weighted versus T2-weighted imaging and myocardial necrosis markers. Tex Heart Inst J 2016; 43(5): 383-391.

Chawla S, Kim S, Wangl S, et al. Diffusion-weighted imaging in head and neck cancers. Future Oncol 2009; 5(7): 959-975.

Soares JM, Marques P, Alves V, et al. A hitchhiker’s guide to diffusion tensor imaging. Front Neurosci 2013; 7: 31.

Yoshida T, Urikura A, Shirata K, et al. Image quality assessment of single-shot turbo spin echo diffusion-weighted imaging with parallel imaging technique: a phantom study. Br J Radiol 2016; 89(1065): 20160512.

Andre JB, Bammer R. Advanced diffusion-weighted magnetic resonance imaging techniques of the human spinal cord. Top Magn Reson Imaging 2010; 21(6): 367-378.

Olman CA, Davachi L, Inati S. Distortion and signal loss in medial temporal lobe. PLoS One 2009; 4(12): e8160.

Thoeny HC, De Keyzer F, King AD. Diffusion-weighted MR Imaging in the Head and Neck. Radiology 2012; 263(1): 19-32.

Stockmann JP, Wald LL. In vivo B0 field shimming methods for MRI at 7T. Neuroimage 2018; 168: 71-87.

Lavdas I, Miquel ME, McRobbie DW, et al. Comparison between diffusion-weighted MRI (DW-MRI) at 1.5 and 3 tesla: a phantom study. J Magn Reson Imaging 2014; 40(3): 682-690.

Porter DA, Heidemann RM. High resolution diffusion-weighted imaging using readout-segmented echo-planar imaging, parallel imaging and a two-dimensional navigator-based reacquisition. Magn Reson Med 2009; 62(2): 468-475.

Filipe JP, Curvo-Semedo L, Casalta-Lopeset J, et al. Diffusion-weighted imaging of the liver: usefulness of ADC values in the differential diagnosis of focal lesions and effect of ROI methods on ADC measurements. MAGMA 2013; 26(3): 303-312.

Zhang H, Xue H, Stemmer A, et al. Integrated shimming improves lesion detection in Whole-Body Diffusion-Weighted examinations of patients with plasma disorder at 3 T. Invest Radiol 2016; 51(5): 297-305.

Barth BK, Cornelius A, Nanzet D, et al. Diffusion-Weighted Imaging of the prostate: Image quality and geometric distortion of Readout-Segmented versus Selective-Excitation Accelerated Acquisitions. Invest Radiol 2015; 50(11): 785-791.

Chen L, Sun P, Hao Q, et al. Diffusion-weighted MRI in the evaluation of the thyroid nodule: Comparison between integrated-shimming EPI and conventional 3D-shimming EPI techniques. Oncotarget 2018; 9(40): 26209-26216.

Darnell D, Truong TK, Song AW. Integrated parallel reception, excitation, and shimming (iPRES) with multiple shim loops per radio-frequency coil element for improved B0 shimming. Magn Reson Med 2017; 77(5): 2077-2086.

Stemmer A, Kiefer B. Combination of integrated slice-specific dynamic shimming and pixel-wise unwarping of residual EPI distortions. Proc Int Soc Magn Reson Med 2015; Abstract #3729.

Kim YJ, Kim SH, Kang BJ, et al. Readout-segmented echo-planar imaging in diffusion-weighted mr imaging in breast cancer: comparison with single-shot echo-planar imaging in image quality. Korean J Radiol 2014; 15(4): 403-410.

Li H, Liu L, Shiet Q, et al. Bladder cancer: detection and image quality compared among iShim, RESOLVE, and ss-EPI diffusion-weighted MR imaging with high b value at 3.0 T MRI. Medicine (Baltimore) 2017; 96(50): e9292.

Gatidis S, Graf H, Weiß J, et al. Diffusion-weighted echo planar MR imaging of the neck at 3 T using integrated shimming: comparison of MR sequence techniques for reducing artifacts caused by magnetic-field inhomogeneities. MAGMA 2017; 30(1): 57-63.

Heroiu Cataloiu AD, Danciu CE, Popescu CR. Multiple cancers of the head and neck. Maedica (Buchar) 2013; 8(1): 80-85.




DOI: http://dx.doi.org/10.36162/hjr.v5i3.338

Refbacks

  • There are currently no refbacks.